SYMBOL CAUSE CORRECTION
Monitor did not detect
pulse rate
Remove the arm cuff. Refer to
Applying the arm cuff” on page
13. Wait 2-3 minutes.
Take another measurement.
Cuff under-inflated
Remove the arm cuff.
Read, “Taking a
Measurement” on page 16.
Wait 2-3 minutes.
Take another measurement.
Cuff over-inflated
Batteries are worn
Replace worn batteries.
Refer to page 8 for
battery installation.
Movement during
measurement
Remove the arm cuff. Wait 2-3
minutes. Take another
measurement. Remain still
during the measurement.
ERROR INDICATORS
21
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 20